Skip to Main Content
Book cover for The ESC Textbook of Cardiovascular Medicine (3 edn) The ESC Textbook of Cardiovascular Medicine (3 edn)

Contents

Update:

Figure 1 has been replaced

New reference added in the Remote Follow-up section

November 28, 2019: This chapter has been re-evaluated and remains ...More

Update:

Figure 1 has been replaced

New reference added in the Remote Follow-up section

November 28, 2019: This chapter has been re-evaluated and remains up-to-date. No changes have been necessary.

Disclaimer
Oxford University Press makes no representation, express or implied, that the drug dosages in this book are correct. Readers must therefore always … More Oxford University Press makes no representation, express or implied, that the drug dosages in this book are correct. Readers must therefore always check the product information and clinical procedures with the most up to date published product information and data sheets provided by the manufacturers and the most recent codes of conduct and safety regulations. The authors and the publishers do not accept responsibility or legal liability for any errors in the text or for the misuse or misapplication of material in this work. Except where otherwise stated, drug dosages and recommendations are for the non-pregnant adult who is not breastfeeding.

Cardiac pacing is the treatment of choice for patients with bradycardia. Several existing algorithms and features facilitate individualization and optimization of device programming on an individual patient basis. Individually based optimal choice of device and pacing mode should be the rule. Correct programming of available pacemaker algorithms and features is a prerequisite for optimal patient treatment.

A pacemaker system for permanent cardiac pacing consists of a pacemaker and one lead (single-chamber pacemaker) or two leads (dual-chamber pacemaker) implanted in the right atrium or right ventricle, or both. The pacemaker contains a battery, the power source, and a pulse generator, the electronic unit controlling the behaviour of the pacemaker (see Chapter 38.17 for more details).

Most commonly, the power source is a constant-voltage lithium–iodine battery that has predictable voltage behaviour over time. The battery voltage remains relatively constant throughout most of its discharge. Expected battery longevity for cardiac pacemakers is 6–12 years. Battery current drain is highly dependent on pacemaker programming. A pacemaker with 6 years of battery longevity under nominal pacing parameters may reach its replacement time at 2 years at one extreme or more than 10 years at the other extreme. The clinical indicators of battery depletion vary between manufacturers and models, the most common being a gradual decrease in telemetered battery voltage. Furthermore, estimated remaining battery longevity is indicated during follow-up and an elective replacement indicator is activated when battery depletion approaches and the pacemaker should be replaced.

Pacemaker leads are bipolar or unipolar. Bipolar systems have both electrodes near the distal end of the lead, the tip electrode acting as the cathode and the more proximal ring electrode as the anode. In a unipolar system, a single electrode (the cathode) is located at the lead tip, and the pacemaker can act as the other electrode (the anode). Bipolar leads reduce the risk of myopotential over-sensing, far-field sensing, crosstalk, and local skeletal muscle stimulation, and they allow programmable switching between bipolar and unipolar configurations. Therefore, bipolar leads should be standard in both the atrium and ventricle. Most bipolar leads have a coaxial design with the inner coil terminating at the tip electrode and the outer coil at the ring electrode.

Active endocardial fixation of pacing leads by screwing a fixed or retractable tip-screw into the atrial or ventricular myocardium is most commonly used. Active fixation reduces the frequency of atrial lead dislodgements and allows for straightforward lead implantation in different locations of both right atrium and ventricle, such as the right atrial septum and right ventricular outflow tract. Today, passive fixation by tines or wings trapped in the trabeculae of the right atrium or ventricle is rarely used. For paediatric patients and patients without venous access to the right side of the heart, epicardial leads are the choice. Modern leads have a small tip surface area with a porous surface, promoting low thresholds, low current drain, and good sensing. Steroid-eluting leads with a small reservoir of glucocorticoids at the lead tip are recommended to diminish the inflammatory reaction at the electrode–endocardial interface and improve acute and chronic pacing thresholds and sensing.

In a pacemaker system, the weak link is the pacing lead, most commonly causing complications. Recently, the concept of leadless pacing was introduced. Two leadless pacing systems were tested in observational feasibility cohort studies and found to meet pre-specified safety and efficacy endpoints.1,2 Using specially designed delivery tools, the leadless pacemaker is introduced via the femoral vein and implanted in the right ventricle where it is anchored either by a helical screw-in fixation electrode1 or by electrically inactive tines.2 So far, leadless pacing is used for single-lead ventricular pacing only. To date, no comparative studies have been conducted that document non-inferiority of leadless pacing to transvenous pacing. Therefore, leadless pacing should be reserved for selected patients in whom traditional transvenous pacing is judged unsuitable or associated with a significantly higher complication rate, such as, for example, no venous access from the upper extremities or haemodialysis with an extremely high infection risk.

Bradycardia caused by atrioventricular (AV) block requires pacing in the ventricle. Dual-chamber pacing is superior to single-lead ventricular pacing in reducing pacemaker syndrome and improving exercise capacity. However, it is well documented that dual-chamber pacing does not prolong survival or reduce morbidity as compared with single-lead ventricular pacing.3,4,5 The current recommendation is to use dual-chamber pacing as the first choice for patients with AV block6,7 (Figure 38.20.1). The choice should be made on an individual basis, taking into consideration complication risk, costs, and individual patient factors. Rate-adaptive single-lead ventricular pacing is a reasonable alternative for inactive, elderly patients with AV block.5 It is well documented that left ventricular ejection fraction is lower with ventricular pacing as compared to intrinsic activation because of the abnormal electrical activation resulting in an abnormal and less synchronous contraction. Therefore, in case of severely reduced left ventricular ejection fraction and clinical heart failure, cardiac resynchronization therapy should be considered at initial implantation if a high percentage of ventricular pacing is expected.8 For patients with persistent atrial fibrillation and AV block, rate-adaptive single-lead ventricular pacing is the mode of choice.6

 Optimal pacing mode and algorithm selection in sinus node dysfunction and atrioventricular block. AF, atrial fibrillation; AV, atrioventricular; AVM, atrioventricular management [i.e. AV delay programming (avoiding values >230 ms) or specific algorithms to avoid/reduce unnecessary ventricular pacing]; CRT, cardiac resynchronization therapy; SND, sinus node dysfunction. (a) (R) indicates that the programming of such a pacing mode is preferred only in the case of chronotropic incompetence. (b) Reasons to avoid two leads include young age and limited venous access. Note: in patients who are candidates for a VVI/VDD pacemaker, a leadless pacemaker may be considered
Figure 38.20.1

Optimal pacing mode and algorithm selection in sinus node dysfunction and atrioventricular block. AF, atrial fibrillation; AV, atrioventricular; AVM, atrioventricular management [i.e. AV delay programming (avoiding values >230 ms) or specific algorithms to avoid/reduce unnecessary ventricular pacing]; CRT, cardiac resynchronization therapy; SND, sinus node dysfunction. (a) (R) indicates that the programming of such a pacing mode is preferred only in the case of chronotropic incompetence. (b) Reasons to avoid two leads include young age and limited venous access. Note: in patients who are candidates for a VVI/VDD pacemaker, a leadless pacemaker may be considered

From
Glikson M, Nielsen JC, Kronborg MB, et al; ESC Scientific Document Group. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021 Sep 14;42(35):3427–3520. doi: 10.1093/eurheartj/ehab364reference
.

For patients with sinus node disease, single-lead ventricular pacing is associated with a higher risk of atrial fibrillation and thromboembolism than single-lead atrial or dual-chamber pacing.4,9 AV block with the need for a ventricular pacing lead occurs in 1.5–1.8% of patients with sinus node disease per year,10,11 and atrial fibrillation is more common with single-lead atrial than with dual-chamber pacing.12,13 Therefore, dual-chamber pacing is the pacing mode of first choice.6 Ventricular pacing should be minimized by AV delay management, either by means of manual optimization of the AV interval or by programming of AV hysteresis (Figure 38.20.1).

All currently available pacemakers possess rate response algorithms intended to imitate physiological increases in heart rate by means of different sensors incorporated into the devices. These sensors convert to pacing rate the strength of an input signal from a surrogate parameter related to physical or emotional exercise.

A range of different sensors have been in use. Currently the most prevalent are the accelerometer (sensing of body acceleration), minute ventilation, and closed loop-based sensors (the latter two are impedance-based between ventricular lead and pacemaker casing). In some devices, sensor blending is used to benefit from the different sensitivity and proportionality in rate response in the individual sensors, for example, an accelerometer sensor is sensitive to initiation and termination of exercise, and a minute ventilation sensor has better proportionality to intensity of exercise.14

While there is good documentation for sensors to deliver rate responses with various and modest sensitivity and proportionality, there is only weak evidence for the clinical benefit.15,16

There are recommendations for using rate response pacing in sick sinus syndrome with clear chronotropic incompetence,17 atrial fibrillation with AV block, and when choosing single-lead ventricular (VVI) pacing in pacing-dependent patients whether in sinus rhythm or atrial fibrillation.17,18

Haemodynamically, the optimal AV delay in pacemaker patients differs between individuals, for most around 125–200 ms at rest.19,20,21,22 In dual-chamber pacemakers, sensed AV delay is typically programmed 20–50 ms shorter than paced AV delay to compensate for atrial sensing taking place after the start of atrial activation. In patients with persistent AV block, the most common practice is to use a rate-adaptive AV delay that shortens with higher heart rate. Small crossover studies indicate better patient well-being and exercise tolerance with rate-adaptive AV delay,21,23 while no larger randomized trials have been conducted.

For patients without persistent AV block, ventricular pacing should be minimized by programming. For this purpose, either an individual optimization of the AV delay or programming of AV hysteresis is recommended.6 AV hysteresis automatically increases the AV delay to search for intrinsic AV conduction thereby promoting intrinsic activation. It is well documented that AV hysteresis can reduce ventricular pacing.24,25 Allowing an AV delay up to 250–270 ms is common. Beyond this length of the AV delay, it is not likely that the benefits of avoiding ventricular pacing exceed the benefits of preserving the AV synchrony.12,13 Haemodynamically, a very prolonged AV delay is disadvantageous, and prolonged AV conduction is associated with increased risk of atrial fibrillation.26

Two manufacturers have developed more advanced algorithms to minimize unnecessary ventricular pacing, the ‘managed ventricular pacing’ algorithm in Medtronic devices, and the ‘SafeR’ algorithm in Sorin devices. Both algorithms act as single-lead atrial pacing modes switching to dual-chamber pacing modes in case of defined higher levels of AV block. These algorithms are the most effective in reducing ventricular pacing.27,28,29,30 As compared with dual-chamber pacing with a very high percentage of ventricular stimulation, managed ventricular pacing reduces atrial fibrillation in patients with sinus node disease.31 However, as compared with dual-chamber pacing programmed with a normal or slightly prolonged AV delay, neither of these two algorithms provide any clinical benefit with respect to preventing atrial fibrillation or other important clinical endpoints.32,33,34 Furthermore, several cases have been published of ventricular fibrillation induced by short–long–short sequences occurring with managed ventricular pacing.35,36,37,38 These algorithms allowing intermittent AV block therefore should not be used for patients with AV block.

Tracking modes in pacemakers (e.g. DDD and VDD) have the inherent risk of tracking atrial tachyarrhythmia to the maximum tracking rate.39 To avoid this, mode switching to a non-tracking mode is necessary when atrial tachyarrhythmia occurs, usually to DDI(R) or VDI(R) (uninterrupted atrial sensing to decide on termination of the arrhythmia). The algorithms governing mode switching differ between device companies.40,41 A prerequisite for reliable mode switching is good atrial sensing (e.g. of low-amplitude atrial fibrillation). The various algorithms differ with regard to degree of programmability, number of fast atrial intervals required, consecutive interval counting, mean atrial interval, and so on, before mode switching occurs.41 The atrial tachycardia detection rate is usually programmable. Basic pacemaker intervals are an important consideration in evaluating mode-switching performance. While atrial tachycardia sensing can occur during the post-ventricular atrial refractory period, the post-ventricular atrial blanking period does not allow atrial sensing. Potentially, this may lead to a lack of mode switching in regular atrial arrhythmias such as atrial flutter,42 where every other atrial event may coincide with the post-ventricular atrial blanking period. For this reason, many devices include an additional mode-switching-related algorithm, atrial flutter response, to reveal underlying atrial flutter in such situations.43 Mode switching should be activated whenever a tracking mode is programmed.17

The most common pacemaker-mediated tachycardia is endless loop tachycardia (ELT) which can occur in dual-chamber pacing modes with atrial tracking.39 The refractory interval post-ventricular atrial refractory period prevents ELT in most circumstances. Commonly, ELT is initiated by premature ventricular contractions, giving rise to retrograde conduction to the atrium beyond the post-ventricular atrial refractory period, leading to tracking of the sensed retrograde atrial event, constituting the loop. Preventive algorithms differ between devices, but usually include extension of the post-ventricular atrial refractory period after premature ventricular contractions and algorithms to terminate the ELT by interrupting the loop (e.g. by withholding one ventricular paced beat, extending the post-ventricular atrial refractory period, or by delivering atrial pacing when atrial sensing and ventricular pacing occurs at maximal tracking rate, or at a programmable ELT rate). Detection algorithms differ between devices (AV interval modulation to detect VA linking indicative of ELT, ELT rate programmability, etc.).

In general, it is recommended to programme these features on when dual-chamber pacing in an atrial tracking mode is operative.

Most contemporary devices include automatic threshold determination as a programmable option for atrial and right ventricular leads. Algorithms differ between companies mainly with regard to whether there is beat-to-beat capture verification or not.39 They depend on detecting the evoked response in the myocardium from the pace stimulus and discerning this from the polarization voltage on the lead tip. The output is adjusted automatically with a programmable safety margin above the measured threshold. Safety in pacing is provided when threshold changes are reliably detected. These automatic features are most beneficial when high pacing percentages or high thresholds are present and increased battery longevity can be anticipated, or when remote monitoring is planned.44,45 Pacing thresholds should still be tested at in-clinic visits to verify correspondence with automatic measured values.

Rate hysteresis allows the rate to go below the nominal lower rate limit. When intrinsic beats are sensed, pacing will occur when reaching the hysteresis rate, while paced beats resume at the programmed lower rate if intrinsic events are not registered. The related feature, night rate, in effect allows a programmable lower rate limit at night-time below the nominal lower rate limit and mimics physiological diurnal rate variation.

These features are recommended on an individual basis. No evidence for beneficial effect is present. By reducing pacing, they can be battery conserving.

Algorithms to terminate or prevent atrial arrhythmias have been applied with equivocal evidence for effect. Atrial overdrive and preference pacing to suppress premature atrial beats supposed to initiate atrial tachyarrhythmia as well as non-competitive atrial pacing to avoid pacing in the atrial vulnerable phase are present in different variations in many devices. While seemingly physiologically rational, the evidence is not strong for clinical effect.46

Active pacing termination features exist in a few devices with the assumption that even more disorganized atrial rhythms (i.e. atrial fibrillation episodes) are initiated by short bursts of more organized atrial arrhythmia amenable to pace termination. The evidence is controversial.32,47,48 However, in the light of ample evidence that automatic focus firing from the pulmonary veins initiates atrial fibrillation episodes, this assumption is controversial, lending more support to the preventive features.

These algorithms should be programmed on an individual basis.17

Modern pacemakers sample a vast amount of diagnostically useful follow-up information39 (for details, refer to Chapter 38.22). Importantly, some of this information is only sampled if the pacemaker is programmed to do so, for example, high ventricular rate episodes, pre-episode electrogram storage, and automatic capture threshold information over time. Information on the performance of some of the algorithms described previously in the ‘Atrioventricular delay programming’ section is included in follow-up data (e.g. percentage of time spent in mode switch, sensor-indicated rate distribution, episodes with ELT, managed ventricular pacing function, and AV interval) giving opportunity for refining the function of these algorithms in the individual patient at follow-up.

Magnetic resonance imaging (MRI) conditional pacemakers contain a MRI safe mode algorithm to be programmed on during MRI, but currently only a minority of implanted systems will be MRI safe. However, most pacing systems can undergo MRI (1.5 T MRI) at minor risk,49 provided that appropriate reprogramming is performed according to management guidelines.17 An important point in relation to programming is whether the patient is pacemaker dependent, that is, whether an asynchronous or an inhibited mode should be programmed during MRI. Features that could elicit inappropriate pacing, such as rate smoothing, trigger pacing, and conducted atrial fibrillation response, must be programmed off. A relative contraindication for MRI is the presence of abandoned leads.50

graphic Remote follow-up is becoming the standard of care in the implantable cardioverter defibrillator population. In pacemaker patients, the cost:benefit ratio is less favourable. Most contemporary pacemakers have the option to be followed in remote monitoring systems. Presently, the choice to do so is based on an individual assessment (e.g. in older non-ambulatory patients or other situations where specific reasons favour this option). According to the most recent ESC GL on cardiac pacing, remote device management is recommended in patients who have difficulties attending in-office follow-up and in patients who have a device component that is recalled or on advisory.51 Remote device management may allow spacing in-office follow-up visits up to 24 months. Automatic threshold measurement and other options should be activated to derive most benefit (see Chapter 38.22 for more details).

1.  Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, Friedman PA, Estes NA 3rd, Ip J, Niazi I, Plunkitt K, Banker R, Porterfield J, Ip JE, Dukkipati SR; LEADLESS II Study Investigators.

Percutaneous implantation of an entirely intracardiac leadless pacemaker.
 
N Engl J Med
 
2015
;373:1125–35.

2.  Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, Narasimhan C, Steinwender C, Brugada J, Lloyd M, Roberts PR, Sagi V, Hummel J, Bongiorni MG, Knops RE, Ellis CR, Gornick CC, Bernabei MA, Laager V, Stromberg K, Williams ER, Hudnall JH, Ritter P.  

A leadless intracardiac transcatheter pacing system.
 
N Engl J Med
 
2016
;374:533–41.

3.  Connolly SJ, Kerr CR, Gent M, Roberts RS, Yusuf S, Gillis AM, Sami MH, Talajic M, Tang AS, Klein GJ, Lau C, Newman DM.  

Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. Canadian Trial of Physiologic Pacing Investigators.
 
N Engl J Med
 
2000
;342:1385–91.

4.  Healey JS, Toff WD, Lamas GA, Andersen HR, Thorpe KE, Ellenbogen KA, Lee KL, Skene AM, Schron EB, Skehan JD, Goldman L, Roberts RS, Camm AJ, Yusuf S, Connolly SJ.  

Cardiovascular outcomes with atrial-based pacing compared with ventricular pacing: meta-analysis of randomized trials, using individual patient data.
 
Circulation
 
2006
;114:11–7.

5.  Toff WD, Camm AJ, Skehan JD, United Kingdom P, Cardiovascular Events Trial I.

Single-chamber versus dual-chamber pacing for high-grade atrioventricular block.
 
N Engl J Med
 
2005
;353:145–55.

6.  Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, Cleland J, Deharo JC, Delgado V, Elliott PM, Gorenek B, Israel CW, Leclercq C, Linde C, Mont L, Padeletti L, Sutton R, Vardas PE; ESC Committee for Practice Guidelines (CPG), Zamorano JL, Achenbach S, Baumgartner H, Bax JJ, Bueno H, Dean V, Deaton C, Erol C, Fagard R, Ferrari R, Hasdai D, Hoes AW, Kirchhof P, Knuuti J, Kolh P, Lancellotti P, Linhart A, Nihoyannopoulos P, Piepoli MF, Ponikowski P, Sirnes PA, Tamargo JL, Tendera M, Torbicki A, Wijns W, Windecker S.  

2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).
 
Eur Heart J
 
2013
;34:2281–329.

7.  Gillis AM, Russo AM, Ellenbogen KA, Swerdlow CD, Olshansky B, Al-Khatib SM, Beshai JF, McComb JM, Nielsen JC, Philpott JM, Shen WK;

Heart Rhythm
Society; American College of Cardiology Foundation.
HRS/ACCF expert consensus statement on pacemaker device and mode selection. Developed in partnership between the Heart Rhythm Society (HRS) and the American College of Cardiology Foundation (ACCF) and in collaboration with the Society of Thoracic Surgeons.
Heart Rhythm
2012
;9:1344–65.

8.  Curtis AB, Worley SJ, Adamson PB, Chung ES, Niazi I, Sherfesee L, Shinn T, Sutton MS; Biventricular versus Right Ventricular Pacing in Heart Failure Patients with Atrioventricular Block (BLOCK HF) Trial Investigators.

Biventricular pacing for atrioventricular block and systolic dysfunction.
 
N Engl J Med
 
2013
;368:1585–93.

9.  Lamas GA, Lee KL, Sweeney MO, Silverman R, Leon A, Yee R, Marinchak RA, Flaker G, Schron E, Orav EJ, Hellkamp AS, Greer S, McAnulty J, Ellenbogen K, Ehlert F, Freedman RA, Estes NA 3rd, Greenspon A, Goldman L.  

Ventricular pacing or dual-chamber pacing for sinus-node dysfunction.
 
N Engl J Med
 
2002
;346:1854–62.

10.  Brandt J, Anderson H, Fahraeus T, Schuller H.  

Natural history of sinus node disease treated with atrial pacing in 213 patients: implications for selection of stimulation mode.
 
J Am Coll Cardiol
 
1992
;20:633–9.

11.  Kirkfeldt RE, Andersen HR, Nielsen JC, DANPACE Investigators.

System upgrade and its complications in patients with a single lead atrial pacemaker: data from the DANPACE trial.
 
Europace
 
2013
;15:1166–73.

12.  Nielsen JC, Thomsen PE, Hojberg S, Moller M, Riahi S, Dalsgaard D, Mortensen LS, Nielsen T, Asklund M, Friis EV, Christensen PD, Simonsen EH, Eriksen UH, Jensen GV, Svendsen JH, Toff WD, Healey JS, Andersen HR.  

Atrial fibrillation in patients with sick sinus syndrome: the association with PQ-interval and percentage of ventricular pacing.
 
Europace
 
2012
;14:682–9.

13.  Nielsen JC, Thomsen PE, Hojberg S, Moller M, Vesterlund T, Dalsgaard D, Mortensen LS, Nielsen T, Asklund M, Friis EV, Christensen PD, Simonsen EH, Eriksen UH, Jensen GV, Svendsen JH, Toff WD, Healey JS, Andersen HR.  

A comparison of single-lead atrial pacing with dual-chamber pacing in sick sinus syndrome.
 
Eur Heart J
 
2011
;32:686–96.

14.  Padeletti L, Pieragnoli P, Di Biase L, Colella A, Landolina M, Moro E, Orazi S, Vicentini A, Maglia G, Pensabene O, Raciti G, Barold SS.  

Is a dual-sensor pacemaker appropriate in patients with sino-atrial disease? Results from the DUSISLOG study.
 
Pacing Clin Electrophysiol
 
2006
;29:34–40.

15.  Lamas GA, Knight JD, Sweeney MO, Mianulli M, Jorapur V, Khalighi K, Cook JR, Silverman R, Rosenthal L, Clapp-Channing N, Lee KL, Mark DB.  

Impact of rate-modulated pacing on quality of life and exercise capacity – evidence from the Advanced Elements of Pacing Randomized Controlled Trial (ADEPT).
 
Heart Rhythm
 
2007
;4:1125–32.

16.  Coman J, Freedman R, Koplan BA, Reeves R, Santucci P, Stolen KQ, Kraus SM, Meyer TE; LIFE Study Results.

A blended sensor restores chronotropic response more favorably than an accelerometer alone in pacemaker patients: the LIFE study results.
 
Pacing Clin Electrophysiol
 
2008
;31:1433–42.

17.  Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, Cleland J, Deharo JC, Delgado V, Elliott PM, Gorenek B, Israel CW, Leclercq C, Linde C, Mont L, Padeletti L, Sutton R, Vardas PE; ESC Committee for Practice Guidelines (CPG), Zamorano JL, Achenbach S, Baumgartner H, Bax JJ, Bueno H, Dean V, Deaton C, Erol C, Fagard R, Ferrari R, Hasdai D, Hoes AW, Kirchhof P, Knuuti J, Kolh P, Lancellotti P, Linhart A, Nihoyannopoulos P, Piepoli MF, Ponikowski P, Sirnes PA, Tamargo JL, Tendera M, Torbicki A, Wijns W, Windecker S.  

2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).
 
Eur Heart J
 
2013
;34:2281–329.

18.  Lau CP, Rushby J, Leigh-Jones M, Tam CYF, Poloniecki J, Ingram A, Sutton R, Camm AJ.  

Symptomatology and quality of life in patients with rate-responsive pacemakers: a double-blind, randomized, crossover study.
 
Clin Cardiol
 
1989
;12:505–12.

19.  Ishikawa T, Sugano T, Sumita S, Kimura K, Kikuchi M, Kosuge M, Kobayashi I, Shigemasa T, Endo T, Usui T, Umemura S.  

Relationship between atrioventricular delay, QT interval and cardiac function in patients with implanted DDD pacemakers.
 
Europace
 
1999
;1:192–6.

20.  Ishikawa T, Sumita S, Kimura K, Kuji N, Nakayama R, Nagura T, Miyazaki N, Tochikubo O, Usui T, Kashiwagi M, Ishii M.  

Critical PQ interval for the appearance of diastolic mitral regurgitation and optimal PQ interval in patients implanted with DDD pacemakers.
 
Pacing Clin Electrophysiol
 
1994
;17:1989–94.

21.  Khairy P, Talajic M, Dominguez M, Tardif JC, Juneau M, Lavoie L, Roy D, Dubuc M.  

Atrioventricular interval optimization and exercise tolerance.
 
Pacing Clin Electrophysiol
 
2001
;24:1534–40.

22.  Ovsyshcher I, Zimlichman R, Katz A, Bondy C, Furman S.  

Measurements of cardiac output by impedance cardiography in pacemaker patients at rest: effects of various atrioventricular delays.
 
J Am Coll Cardiol
 
1993
;21:761–7.

23.  Sulke AN, Chambers JB, Sowton E.  

The effect of atrio-ventricular delay programming in patients with DDDR pacemakers.
 
Eur Heart J
 
1992
;13:464–72.

24.  Bauer A, Vermeulen J, Toivonen L, Voitk J, Barr C, Peytchev P.  

Minimizing right ventricular pacing in pacemaker patients with intact and compromised atrioventricular conduction: results from the EVITA Trial.
 
Herzschrittmacherther Elektrophysiol
 
2015
;26:359–66.

25.  Kolb C, Schmidt R, Dietl JU, Weyerbrock S, Morgenstern M, Fleckenstein M, Beier T, Von Bary C, Mackes KG, Widmaier J, Kreuzer J, Semmler V, Zrenner B.  

Reduction of right ventricular pacing with advanced atrioventricular search hysteresis: results of the PREVENT study.
 
Pacing Clin Electrophysiol
 
2011
;34:975–83.

26.  Kwok CS, Rashid M, Beynon R, Barker D, Patwala A, Morley-Davies A, Satchithananda D, Nolan J, Myint PK, Buchan I, Loke YK, Mamas MA.  

Prolonged PR interval, first-degree heart block and adverse cardiovascular outcomes: a systematic review and meta-analysis.
 
Heart
 
2016
;102:672–80.

27.  Davy JM, Hoffmann E, Frey A, Jocham K, Rossi S, Dupuis JM, Frabetti L, Ducloux P, Prades E, Jauvert G.  

Near elimination of ventricular pacing in SafeR mode compared to DDD modes: a randomized study of 422 patients.
 
Pacing Clin Electrophysiol
 
2012
;35:392–402.

28.  Frohlig G, Gras D, Victor J, Mabo P, Galley D, Savouré A, Jauvert G, Defaye P, Ducloux P, Amblard A.  

Use of a new cardiac pacing mode designed to eliminate unnecessary ventricular pacing.
 
Europace
 
2006
;8:96–101.

29.  Gillis AM, Purerfellner H, Israel CW, Sunthorn H, Kacet S, Anelli-Monti M, Tang F, Young M, Boriani G.  

Reducing unnecessary right ventricular pacing with the managed ventricular pacing mode in patients with sinus node disease and AV block.
 
Pacing Clin Electrophysiol
 
2006
;29:697–705.

30.  Murakami Y, Tsuboi N, Inden Y, Yoshida Y, Murohara T, Ihara Z, Takami M.  

Difference in percentage of ventricular pacing between two algorithms for minimizing ventricular pacing: results of the IDEAL RVP (Identify the Best Algorithm for Reducing Unnecessary Right Ventricular Pacing) study.
 
Europace
 
2010
;12:96–102.

31.  Sweeney MO, Bank AJ, Nsah E, Koullick M, Zeng QC, Hettrick D, Sheldon T, Lamas GA.  

Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease.
 
N Engl J Med
 
2007
;357:1000–8.

32.  Boriani G, Tukkie R, Manolis AS, Mont L, Purerfellner H, Santini M, Inama G, Serra P, de Sousa J, Botto GL, Mangoni L, Grammatico A, Padeletti L.  

Atrial antitachycardia pacing and managed ventricular pacing in bradycardia patients with paroxysmal or persistent atrial tachyarrhythmias: the MINERVA randomized multicentre international trial.
 
Eur Heart J
 
2014
;35:2352–62.

33.  Botto GL, Ricci RP, Benezet JM, Nielsen JC, De Roy L, Piot O, Quesada A, Quaglione R, Vaccari D, Garutti C, Vainer L, Kozák M.  

Managed ventricular pacing compared with conventional dual-chamber pacing for elective replacement in chronically paced patients: results of the Prefer for Elective Replacement Managed Ventricular Pacing randomized study.
 
Heart Rhythm
 
2014
;11:992–1000.

34.  Stockburger M, Boveda S, Moreno J, Da Costa A, Hatala R, Brachmann J, Butter C, Garcia Seara J, Rolando M, Defaye P.  

Long-term clinical effects of ventricular pacing reduction with a changeover mode to minimize ventricular pacing in a general pacemaker population.
 
Eur Heart J
 
2015
;36:151–7.

35.  Mansour F, Khairy P.  

Electrical storm due to managed ventricular pacing.
 
Heart Rhythm
 
2012
;9:842–3.

36.  Sekita G, Hayashi H, Nakazato Y, Daida H.  

Ventricular fibrillation induced by short-long-short sequence during managed ventricular pacing.
 
J Cardiovasc Electrophysiol
 
2011
;22:1181.

37.  van Mechelen R, Schoonderwoerd R.  

Risk of managed ventricular pacing in a patient with heart block.
 
Heart Rhythm
 
2006
;3:1384–5.

38.  Vavasis C, Slotwiner DJ, Goldner BG, Cheung JW.  

Frequent recurrent polymorphic ventricular tachycardia during sleep due to managed ventricular pacing.
 
Pacing Clin Electrophysiol
 
2010
;33:641–4.

39.  Ellenbogen KA, Kay GN, Lau CK, Wilkoff BL (eds).

Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy.
4th ed. Philadelphia, PA: Elsevier;
2011
.

40.  Marshall HJ, Kay GN, Hess M, Plumb VJ, Bubien RS, Hummel J, Dawson D, Markewitz T, Gammage MD.  

Mode switching in dual chamber pacemakers: effect of onset criteria on arrhythmia-related symptoms.
 
Europace
 
1999
;1:49–54.

41.  Lau CP, Leung SK, Tse HF, Barold SS.  

Automatic mode switching of implantable pacemakers: II. Clinical performance of current algorithms and their programming.
 
Pacing Clin Electrophysiol
 
2002
;25:1094–113.

42.  Israel CW, Barold SS.  

Failure of atrial flutter detection by a pacemaker with a dedicated atrial flutter detection algorithm.
 
Pacing Clin Electrophysiol
 
2002
;25:1274–7.

43.  Goethals M, Timmermans W, Geelen P, Backers J, Brugada P.  

Mode switching failure during atrial flutter: the ‘2:1 lock-in’ phenomenon.
 
Europace
 
2003
;5:95–102.

44.  Koplan BA, Gilligan DM, Nguyen LS, Lau TK, Thackeray LM, Berg KC, ULTRA Study Investigators.

A randomized trial of the effect of automated ventricular capture on device longevity and threshold measurement in pacemaker patients.
 
Pacing Clin Electrophysiol
 
2008
;31:1467–74.

45.  Biffi M, Bertini M, Saporito D, Ziacchi M, Martignani C, Diemberger I, Boriani G.  

Actual pacemaker longevity: the benefit of stimulation by automatic capture verification.
 
Pacing Clin Electrophysiol
 
2010
;33:873–1.

46.  Israel CW, Grönefeld G, Ehrlich JR, Li YG, Hohnloser SH.  

Prevention of immediate reinitiation of atrial tachyarrhythmias by high-rate overdrive pacing: results from a prospective randomized trial.
 
J Cardiovasc Electrophysiol
 
2003
;14:954–9.

47.  Lee MA, Weachter R, Pollak S, Kremers MS, Naik AM, Silverman R, Tuzi J, Wang W, Johnson LJ, Euler DE.  

The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency: results of a randomized trial in patients with bradycardia and atrial tachyarrhythmias.
 
J Am Coll Cardiol
 
2003
;41:1926–32.

48.  Gillis AM, Koehler J, Morck M, Mehra R, Hettrick DA.  

High atrial antitachycardia pacing therapy efficacy is associated with a reduction in atrial tachyarrhythmia burden in a subset of patients with sinus node dysfunction and paroxysmal atrial fibrillation.
 
Heart Rhythm
 
2005
;2:791–6.

49.  Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR.  

A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.
 
Ann Intern Med
 
2011
;155:415–24.

50.  Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, Candinas R, Boesiger P.  

In vivo heating of pacemaker leads during magnetic resonance imaging.
 
Eur Heart J
 
2005
;26:376–83.

51.  Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, Barrabés JA, Boriani G, Braunschweig F, Brignole M, Burri H, Coats AJS, Deharo J-C, Delgado V, Diller G-P, Israel CW, Keren A, Knops RE, Kotecha D, Leclercq C, Merkely B, Starck C, Thylén I, Tolosana JM; ESC Scientific Document Group.

2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy.
 
Eur Heart J
 
2021
;42(35):3427–520.

Boriani G, Tukkie R, Manolis AS, Mont L, Purerfellner H, Santini M, Inama G, Serra P, de Sousa J, Botto GL, Mangoni L, Grammatico A, Padeletti L.  

Atrial antitachycardia pacing and managed ventricular pacing in bradycardia patients with paroxysmal or persistent atrial tachyarrhythmias: the MINERVA randomized multicentre international trial.
 
Eur Heart J
 
2014
;35:2352–62.

Botto GL, Ricci RP, Benezet JM, Nielsen JC, De Roy L, Piot O, Quesada A, Quaglione R, Vaccari D, Garutti C, Vainer L, Kozák M.  

Managed ventricular pacing compared with conventional dual-chamber pacing for elective replacement in chronically paced patients: results of the Prefer for Elective Replacement Managed Ventricular Pacing randomized study.
 
Heart Rhythm
 
2014
;11:992–1000.

Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, Cleland J, Deharo JC, Delgado V, Elliott PM, Gorenek B, Israel CW, Leclercq C, Linde C, Mont L, Padeletti L, Sutton R, Vardas PE; ESC Committee for Practice Guidelines (CPG), Zamorano JL, Achenbach S, Baumgartner H, Bax JJ, Bueno H, Dean V, Deaton C, Erol C, Fagard R, Ferrari R, Hasdai D, Hoes AW, Kirchhof P, Knuuti J, Kolh P, Lancellotti P, Linhart A, Nihoyannopoulos P, Piepoli MF, Ponikowski P, Sirnes PA, Tamargo JL, Tendera M, Torbicki A, Wijns W, Windecker S.  

2013
ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).
Eur Heart J
2013;34:2281–329.

Curtis AB, Worley SJ, Adamson PB, Chung ES, Niazi I, Sherfesee L, Shinn T, Sutton MS; Biventricular versus Right Ventricular Pacing in Heart Failure Patients with Atrioventricular Block (BLOCK HF) Trial Investigators.

Biventricular pacing for atrioventricular block and systolic dysfunction.
 
N Engl J Med
 
2013
;368:1585–93.

Ellenbogen KA, Kay GN, Lau CK, Wilkoff BL (eds).

Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy.
4th ed. Philadelphia, PA: Elsevier;
2011
.

Gillis AM, Russo AM, Ellenbogen KA, Swerdlow CD, Olshansky B, Al-Khatib SM, Beshai JF, McComb JM, Nielsen JC, Philpott JM, Shen WK;

Heart Rhythm
Society; American College of Cardiology Foundation.
HRS/ACCF expert consensus statement on pacemaker device and mode selection. Developed in partnership between the Heart Rhythm Society (HRS) and the American College of Cardiology Foundation (ACCF) and in collaboration with the Society of Thoracic Surgeons.
Heart Rhythm
2012
;9:1344–65.

Healey JS, Toff WD, Lamas GA, Andersen HR, Thorpe KE, Ellenbogen KA, Lee KL, Skene AM, Schron EB, Skehan JD, Goldman L, Roberts RS, Camm AJ, Yusuf S, Connolly SJ.  

Cardiovascular outcomes with atrial-based pacing compared with ventricular pacing: meta-analysis of randomized trials, using individual patient data.
 
Circulation
 
2006
;114:11–7.

Nielsen JC, Thomsen PE, Hojberg S, Moller M, Vesterlund T, Dalsgaard D, Mortensen LS, Nielsen T, Asklund M, Friis EV, Christensen PD, Simonsen EH, Eriksen UH, Jensen GV, Svendsen JH, Toff WD, Healey JS, Andersen HR.  

A comparison of single-lead atrial pacing with dual-chamber pacing in sick sinus syndrome.
 
Eur Heart J
 
2011
; 32:686–96.

Stockburger M, Boveda S, Moreno J, Da Costa A, Hatala R, Brachmann J, Butter C, Garcia Seara J, Rolando M, Defaye P.  

Long-term clinical effects of ventricular pacing reduction with a changeover mode to minimize ventricular pacing in a general pacemaker population.
 
Eur Heart J
 
2015
;36:151–7.

Close
This Feature Is Available To Subscribers Only

Sign In or Create an Account

Close

This PDF is available to Subscribers Only

View Article Abstract & Purchase Options

For full access to this pdf, sign in to an existing account, or purchase an annual subscription.

Close